AVAPS™
Features
With Average Volume Assured Pressure Support (AVAPS™), inspiratory pressure automatically changes to ensure the set tidal volume.
AVAPS™
- Adapts to changing patient needs
- Improves patient's ventilation efficacy and comfort
- Increases safety by delivering an average tidal volume
- Makes the titration process easier
Who potentially benefits from AVAPS™ technology?
- Restrictive patients benefit because AVAPS™ provides the comfort and leak compensation of a pressure mode and the safety of a guaranteed volume.
- Obese hypoventilation patients benefit because the algorithm compensates for changes in body position and maintains ventilation.
- COPD patients benefit because AVAPS™ achieves a combination of ventilation comfort and efficiency without compromise. The algorithm applies the right pressure at the right time.
 |
| IPAP min |
From EPAP to IPAP max |
| IPAP max |
From IPAP min to 30 cm H2O |
| Target Vte |
From 200 ml to 1500 ml |
| Alarm |
Low Vte |
|
Suggested Protocol*
- Set target tidal volume to 110% of displayed patient tidal volume in the S/T mode or 8 cc per kg of ideal body weight.
- IPAP Max = 25-30 cm H20
IPAP Min = EPAP + 4 cm H2O
*Respironics does not provide medical advice. This protocol is a suggested protocol only and is not intended to replace or modify your institution's established protocols.